Go to the casebook version optimized for the mobile experience
The patient was implanted for primary prevention of obstructive hypertrophic cardiomyopathy (resting gradient: 80 mmHg). He is asymptomatic, but note the sudden death of two of his brothers, one of whom also had OHCM, with ventricular tachycardia bursts, a genetic mutation, and significant septal fibrosis at MRI. The defibrillator is dual-chamber given that the patient is bradycardic under beta-blocker therapy, and in the hope that the DDD pacing from the RV with complete RV capture while maintaining atrial systole will decrease the obstruction in the long term.
A follow-up control was conducted on day 1, and deemed satisfactory.

Your turn
Are you satisfied with the follow-up control ?
- No, the shock impedance is too low
- Yes, completely
- No, ventricular sensing is 2.2 mV
- No, a shock is recorded, which is why a shock impedance value is provided
- No, I am not satisfied with the impedance values of the shock electrodes
However, two weeks later, a telemedicine alert was transmitted.
This is when the patient comes into clinic after the alert
→ Figure 2
Your turn
What is your opinion (1) ?
- A nonsustained ventricular fibrillation episode occurred
- The atrial lead is fractured
- The ventricular pacing/sensing lead is fractured
- The shock electrode conductor is fractured
- Need to reoperate urgently
Additional information from the alert screen
→ Figure 3
Lead sensing and impedance trends
→ Figure 4
Ventricular signal amplitudes
→ Figure 5
EGM saved into the memory of the device as «VF»
→ Figure 6
Your turn
What is your opinion (2) ?
- The drop in ventricular impedance is consistent with the increase in atrial impedance and the drop in P wave amplitude
- Both leads are damaged
- The device does not sense R-waves since only 152 R waves have been detected
- The VF episode corresponds to RV lead fracture signals
- If the artifacts persist, the risk is an inappropriate shock
Episode recorded during the emergency follow-up, after manipulation of the defibrillator case through the skin
→ Figure 7
Interpretation
- At the first follow-up control, 24 hours after defibrillator placement, ventricular sensing is low at 2.2 mV. This parameter must be re-assessed after a few days, although if the detection value remains very low or decreases again, it will be necessary to discuss the repositioning of the ventricular lead. A minimum value of 5 mV is allowed for the amplitude of the R wave at the time of lead placement.
- The teletransmission initiated an alert due to abnormalities in measured lead impedances. The atrial impedance exceeded the upper limit of 3000 Ohm, signalling a fractured conductor, and the ventricular lead exceeded the lower limit of 200 Ohm, signalling an internal insulation fracture.
- These anomalies occurred at the same time on both leads and suggest a mechanical aggression on the two leads, probably related to a traumatic lead fixation given their occurrence soon after the installation of the device.
- The problem cannot be a connection fault at the level of the defibrillator case, since in this instance, the impedances of the two leads would be beyond the maximum values.
- Only 152 R-waves were sensed since the defibrillator is constantly pacing. R-waves in VF and VT are fracture artifacts and can have a very large amplitude.
- These artifacts occurring in runs lead to false VF diagnoses, which induces capacitor charges.
- It is necessary to reoperate urgently and change the two leads. In the meantime, the therapies are inactivated.
- The handling of the material through the skin « finished off » the leads. The last electrogram no longer reveals any physiological signal, and the artifacts saturate the atrial channel.
Correct answers: Follow-up : 3 Opinion n°1 : 2 and 5 Opinion n°2 : 1, 2, 4 and 5