New device, dual R-wave counting and SecureSense™ algorithm

Patient

Go to the casebook version optimized for the mobile experience

Man implanted with an Abbott Gallant CRT defibrillator; episodes of unsustained V oversensing recorded in memory; SecureSense™ algorithm ON (VD coil to housing)

New device, dual R-wave counting and SecureSense™ algorithm: Abbott ICD tracing (EGM)


Loading Viewer…


Trace

  1. Slow VT (atrioventricular dissociation with V>A)
  2. double counting of the R wave on the bipolar channel but not on the discrimination channel
  3. The noise counter is at 10; diagnosis of unsupported V oversensing (SNS)

Comments


episode of non-sustained ventricular oversensing

Points to remember

  • This patient had episodes of slow VT with very wide QRS complexes, explaining the double counting of the R wave at the end of the ventricular refractory period.
  • This type of oversensing can lead to aggressive therapies (repeated electrical shocks
  • ) following a false diagnosis of VF (slow VT at 120 bpm with double counting and tachycardia detected at 240 bpm in the VF zone).
  • The SecureSense™ algorithm enables early diagnosis even when oversensing is intermittent and would inhibit therapies in cases where a tachycardia counter is completed