Skip to main content
MICROPORT ACADEMY CRM
DAI Boston Scientific
BIOMONITOR IIIm BIOTRONIK

Pacing & Defibrillation

Une formation unique en rythmologie, + de 800 cas cliniques répertoriés, du basique à l’expert.

Quadripolar LV lead threshold test

Patient - EN

62-year-old man implanted with a triple chamber defibrillator Viva Quad XT CRT-D connected to a quadripolar Medtronic 4298 LV lead for ischemic cardiomyopathy with left bundle branch block; the implantation procedure was difficult with few lateral veins of small caliber and a large inferior vein draining the lateral wall; presence of high threshold values in the apical portion of the lateral wall and phrenic nerve stimulation; positioning of a quadripolar lead; LV threshold performed with the different pacing configurations (DDD mode 90 bpm, amplitude decrement, pulse width 0.4 ms).

Graph and trace
  1. LV threshold in LV1 – RV coil configuration;
  2. loss of LV capture (threshold 0.5 V/0.4 ms); absence of phrenic nerve stimulation;
  3. LV threshold in LV1 – LV2 configuration;
  4. loss of LV capture (threshold 0.75 V/0.4 ms); absence of phrenic nerve stimulation;
  5. LV threshold in LV1 – LV3 configuration;
  6. loss of LV capture (threshold 0.5 V/0.4 ms); absence of phrenic nerve stimulation;
  7. LV threshold in LV1 – LV4 configuration;
  8. loss of LV capture (threshold 1 V/0.4 ms); presence of phrenic nerve stimulation (phrenic nerve threshold 2 V/0.4 ms);
  9. LV threshold in LV2 – RV coil configuration;
  10.  loss of LV capture (threshold 2.25 V/0.4 ms); absence of phrenic nerve stimulation;
  11.  LV threshold in LV2 – LV1 configuration;
  12.  loss of LV capture (threshold 2.75 V/0.4 ms); absence of phrenic nerve stimulation;
  13.  LV threshold in LV2 – LV3 configuration;
  14.  loss of LV capture (threshold 6 V/0.4 ms); absence of phrenic nerve stimulation;
  15.  LV threshold in LV2 – LV4 configuration;
  16.  loss of LV capture (threshold 6 V/0.4 ms); presence of phrenic nerve stimulation up to the threshold value;
  17.  LV threshold in LV3 – RV coil configuration;
  18.  loss of LV capture (threshold 4.75 V/0.4 ms); absence of phrenic nerve stimulation;
  19.  LV threshold in LV3 – LV1 configuration;
  20.  loss of LV capture (threshold 2.75 V/0.4 ms); absence of phrenic nerve stimulation;
  21.  LV threshold in LV3 – LV2 configuration;
  22.  loss of LV capture (threshold 6 V/1.5ms); absence of phrenic nerve stimulation;
  23.  LV threshold in LV3 – LV4 configuration;
  24.  loss of LV capture (threshold 6 V/1.5 ms); presence of phrenic nerve stimulation up to the threshold value;
  25.  LV threshold in LV4 – RV coil configuration;
  26.  loss of LV capture (threshold 8 V/0.4 ms); presence of phrenic nerve stimulation up to the threshold value;
  27.  LV threshold in LV4 – LV1 configuration;
  28.  loss of LV capture (threshold 2.75 V/0.4 ms); presence of phrenic nerve stimulation up to the threshold value;
  29.  LV threshold in LV4 – LV2 configuration;
  30.  loss of LV capture (threshold 8 V/1.5 ms); presence of phrenic nerve stimulation up to the threshold value;
  31.  LV threshold in LV4 – LV3 configuration;
  32.  loss of LV capture (threshold 8 V/1.5 ms); presence of phrenic nerve stimulation up to the threshold value.
Comments

These tracings show the various programming options when a left ventricular quadripolar lead is implanted. Given the multitude of possible configurations, the choice of programming can be complex even if there are a certain number of known fundamentals. In a first step, it appears logical to avoid all configurations associated with phrenic stimulation which is often poorly supported by the patient. Similarly, given the substantial variation in pacing thresholds according to the electrode used, e.g. from a very satisfactory value (<1 volt for 0.4 ms) to values ​​nearing the maximum capacities of the device, it would thus seem advantageous not to select a configuration leading to premature wear of the batteries. A certain number of configurations correspond to these first two criteria (absence of phrenic stimulation, necessity of small pacing amplitude). The hemodynamic criterion could then be taken into account for deciding which configuration to choose. The current literature is rich in terms of publications assessing the impact of the left ventricular pacing site on the degree of response to resynchronization. Unfortunately, the results are controversial and suggest the existence of considerable inter-individual variations without the possibility of defining a same universal optimal site. It seems, however, that an apical position should be avoided, which allows eliminating a certain number of configurations. The 4 electrodes on this type of lead are relatively interspaced: in this patient, the distal electrode (LV1) was positioned in a lateral region whereas the proximal electrode (LV4) was positioned opposite a more apical region (lead positioned in a lower vein continuing into a more basal lateral vein). The electrocardiographic variations are relatively invisible on the lead used for the threshold assessment. In contrast, the 12-lead electrocardiogram found significant changes in the paced QRS pattern according to the chosen configuration. Since the electrical activation is different, it would seem plausible to also envision a different degree of response according to the selected configuration. However, given the absence of a validated method to compare the degree of response to be expected according to the pacing site chosen, it remains very difficult to choose a programming based on a possible superior clinical benefit. The possibility of choosing a pacing site without phrenic stimulation with an acceptable pacing threshold while reducing the risk of dislodgement (possibility of pushing the leads more distally in the vein) probably constitutes the main advantage provided by the quadripolar leads. This likely increases the percentage of patients with effective left ventricular pacing. The new devices allow dual-LV pacing with 2 simultaneously-active ventricular lead electrodes. The clinical value of this type of programming remains to be demonstrated.

In this patient, the selected configuration was LV1/RV coil which allowed obtaining a satisfactory threshold and absence of phrenic stimulation.

NID old - EN
3085
Haut de page
Tracé
4
Constructeur
Medtronic
Prothèse
CRT
Chapitre
CRT, Left ventricular pacing