Right atrial automatic threshold (RAAT):
- required bipolar atrial lead
- threshold tests are scheduled every 21 hours
- evaluation of atrial capture is based on the presence of an atrial evoked response
- amplitude is automatically adjusted to 2X the highest of the last 7 days with successful threshold tests at a non-programmable pulse width of 0.4 ms
- the output is limited between 2.0 V and 5.0 V
Determining the RA pacing threshold
- evaluation of atrial capture is based on the presence of an atrial evoked response
- atrial pacing rate is based on the lower rate limit or the sensor rate and be increased by 10 bpm to ensure atrial pacing
- voltage decrements every 3 captured beats
- looks for 2 Loss of Capture (LOC) beats out of 4-beat moving window
- the second LOC in a 4-beat moving window is defined as “Confirmed LOC” (C-LOC)
- the threshold is defined as one voltage step higher than where C-LOC occurs
Adjusting the RA pacing output
- no beat to beat verification of atrial capture
- safety margin: amplitude 2X (non programmable) the highest of the last 7 measured thresholds
- non programmable pulse width: 0.4 ms
- maximal amplitude: 5.0 V
- minimal amplitude: 2.0 V