Search AV + and risk of PMT
Tracing
Manufacturer Boston Scientific
Device PM
Field AV delays, refractory periods management of tachycardia
N° 4
Patient
- 78-year-old man implanted
- Vitalio pacemaker (Boston Scientific)
- Rythmiq and Search AV +
- Episodes of Rythmiq and episodes of PMT
Graph and trace
- After 32 cycles, search for intrinsic activation with prolongation of the AV delay (Search AV +)
- Premature atrial contraction followed by ventricular pacing after a long AV delay,
retrograde conduction and PMT - After 16 consecutive cycles VPMT-AS, prolongation of the PVARP and interruption of the PMT
Comments
- During the search for intrinsic conduction (Search AV +), the 1:1 physiological atrioventricular activation is preserved (no dropped P wave)
- There is an increased risk of pacemaker-mediated tachycardia (temporary prolongation of the AV delay)
- In this patient, the association between retrograde conduction, premature atrial contraction and temporary prolongation of the AV delay is responsible for the PMT episodes recorded in the device memory
- The risk of PMT is nil when the pacemaker is operating in AAI mode with VVI backup
- The risk of PMT is very weak at the time of transition from ADI/VVI backup to DDD mode, the device systematically prolonging the PVARP on one cycle at the time of switch to DDD mode
- The risk of PMT is greater during the search for intrinsic conduction (Search AV +), the temporary prolongation of the AV delay favoring the occurrence of retrograde conduction
Other articles that may be of interest to you
EGM recordings