VIP algorithm
Tracing
Manufacturer Abbott
Device PM
Field Pacing Modes
N° 2
Patient
- 74-year-old man
- AccentTM DR pacemaker (Abbott) for paroxysmal atrioventricular block
- Programming of the VIP algorithm
Graph and trace
- Atrial and ventricular pacing (programmed paced-AV delay 250 ms)
- VIP algorithm; prolongation of the AV delay by 150 ms (programmable) for 2 cycles
- Absence of intrinsic ventricular activity and ventricular pacing at the end of the prolonged AV delay (400 ms)
- Unsuccessful search and return to DDD mode pacing at the programmed AV delay
Comments
- This tracing illustrates the features of the VIP algorithm and its operation in a patient with atrioventricular conduction disorder
- 3 parameters are programmable
- The VIP™ Extension which corresponds to the extension of the AV delay applied to the programmed AV delay (paced or sensed) to search for intrinsic conduction; the total value of the AV delay cannot exceed 450 ms
- The search interval corresponding to the time interval between each intrinsic conduction search which can be programmed to 30 seconds, 1, 3, 5, 10 or 30 minutes; increasing this value avoids iterative searches in a patient with permanent atrioventricular block; decreasing the value avoids unnecessary right ventricular pacing in a patient with short-term paroxysmal conduction disorder
- The number of consecutive cycles the device extends the AV delay to search for the occurrence of intrinsic conduction is programmable to 1, 2 or 3 cycles
- It is possible to view the atrioventricular conduction histogram in the pacemaker's memory
- It is possible to program a specific remote monitoring alert when the percentage of ventricular pacing exceeds a programmable threshold