This 67-year-old man presenting with dilated cardiomyopathy and left bundle branch block received a Unify Assura™ triple chamber ICD. Episodes of non-sustained RV oversensing were stored in the device memories.
This recipient of a cardiac resynchronization system responded imperfectly to treatment, with a percentage of stimulation limited to 70%. The episodes recorded revealed non-sustained RV oversensing, allowing the diagnosis of post-pacing T wave oversensing, explaining the intermittent loss of biventricular stimulation with alternans between stimulated and sensed ventricle. While the SecureSense™ algorithm diagnoses, it does not eliminate T wave oversensing and its stimulation consequences (inhibition of biventricular stimulation after sensing of the T wave). One of the advantages of the Abbott devices is the ability of programming differently the post-ventricular sensing and post-ventricular pacing settings, which require different adjustments. The change in post-pacing sensitivity enabled the correction of oversensing and the recovery of nearly 100% biventricular stimulation, which is mandatory for an optimal response to cardiac resynchronization.