This 65-year-old man received a Abbott Ellipse™ VR defibrillator; an episode of noise on the ventricular lead was stored in the memories.
Tracing 5A
Patient
This 69-year-old man received a Abbott Unify Assura™ triple chamber ICD for management of a dilated cardiomyopathy with non-specific bundle branch block; episode of VF diagnosed and stored in memory.
Tracing 5B
Oversensing of atrial depolarization (P wave) by the RV lead is rare and is observed mainly in recipients of integrated bipolar leads. In patients in sinus rhythm, the RV lead senses both the atrial and ventricular depolarization since the PR interval is longer than the post-sensed ventricular blanking period. In patients presenting with complete AV block, P wave oversensing may inhibit ventricular pacing and cause asystole. Likewise, oversensing of atrial depolarization during flutter or atrial tachycardia may cause both inappropriate therapies and asystole if the patient is pacemaker-dependent.
Oversensing of the atrial signal occurs preferentially in 2 circumstances: 1) in case of RV lead dislodgement at the AV junction (coincides with a fall in R wave amplitude), or 2) when an integrated bipolar lead is implanted near the tricuspid ring, with the distal coil straddling the valve (coinciding with a preserved R wave amplitude). This observation is more likely in patients whose cardiac chambers are small, such as children or patients presenting with hypertrophic cardiomyopathy.
Atrial oversensing may also occur under less common circumstances:
While, as discussed previously, the SecureSense™ algorithm can virtually eliminate the therapies triggered by T wave oversensing, it is less effective against P wave oversensing. If the discrimination channel is set between the coil and the can, oversensing on this channel is more likely (coil in the atrium if the lead is placed near the tricuspid valve) than on the bipolar channel (both electrodes in the ventricle). These 2 tracings illustrate 2 cases of figure:
By changing the discrimination channel to program the detection between the distal electrode and the pulse generator (tip to can), the risk of oversensing on this channel could be lowered, though would not eliminate oversensing on the bipolar channel. In both cases, repositioning of the high-voltage lead must be considered, while verifying during the implant that the distal coil is indeed completely contained in the RV chamber.