This 82-year-old man underwent implantation of a Boston Science Cognis triple chamber defibrillator in the context of severe ischemic cardiomyopathy with left bundle branch block; he was seen in consultation after a shock perceived during effort, while gardening; an alert was delivered by remote monitoring (Latitude).
ng
This tracing is characteristic of a lead fracture; oversensing of very rapid signals is present, with very short cycles, approaching the ventricular refractory period, of variable morphology and intermittent throughout the cardiac cycle. The amplitude of some signals is very high, saturating the amplifiers. These fracture potentials are highly suggestive of lead fracture with probable rupture of the conductor instead of an insulation breakdow