A 47-year-old man presenting with hypertrophic cardiomyopathy and complete atrioventricular block after septal embolization, received a dual chamber Atlas ICD for secondary prevention of ventricular tachycardia (VT).
Main device programming (during the induction procedure)
During the induction test, ventricular sensing is usually set at the highest possible value (1 mV) in order to place the system in adverse sensing conditions and render the test as insensitive as possible. A proper detection of VF at this high value suggests the presence of a wide safety margin with the system programmed at the nominal value of 0.3 mV on the long term. In this patient and for this test, however, the sensitivity was programmed at 0.3 mV, and flaws in ventricular sensing were noted during the induction. Induction tracings must be meticulously scrutinised in order to detect these sensing failures, which may be subtle and challenging to diagnose, as they delay the delivery of shocks only slightly. Sensing failures are facilitated by wide variations in the signal amplitude, which mislead the automatic control of the sensitivity. The failure of sensing low-amplitude VF EGM occurs after the occurrence of a signal of greater amplitude. In this patient, the ventricular sensitivity programmed for the test was 0.3 mV, and ventricular sensing during sinus rhythm, measured at 4.1 mV, was moderately reliable. The decision was made, consequently, to relocate the lead from the right ventricular (RV) apex toward the high septum, where sensing was reliable. A repeat induction procedure confirmed that VF was properly detected and defibrillation was successful.