Patient implanted with a single-chamber defibrillator (Visia AF XT VR) for congenital long QT syndrome; episode of syncope.
The graph shows a sudden acceleration of the ventricular rate with irregular cycles detected in the VF zone; the initial VF counter is completed (Detection) but no therapy is delivered following a probable spontaneous reduction.
This is probably a very rapid and irregular ventricular arrhythmia; the shock canal shows the polymorphic character with a torsadiform appearance.
The initial counter is set to 30/40.
The capacitors are charged when the initial VF counter is completed; during charging, the arrhythmia stops spontaneously and after 4 consecutive slow cycles, the therapy is abandoned; note that the shock channel makes it possible to recover a very elongated QT aspect.