56-year old male patient with a history of severe ischemic cardiomyopathy with left bundle branch block was implanted with a triple-chamber Boston Scientific Autogen CRT-D. Interrogation revealed an episode of mode switch (RTA).
Cross-talk with alternans between 2 different signal morphologies and amplitudes. An alternans between cycles is also seen (the first at approximately 220ms corresponds with oversensing while the second is at approximately 350 ms).
The atrial blanking post ventricular stimulation is the preferred parameter to regulate to correct this problem. The blanking may be programmed ‘SMART’ or at a fixed value. Boston Scientific devices have a different atrial blanking for post-stimulation and for post-detection.
Programming SMART aims to reduce blanking durations while reducing the risk of cross-talk, disabling the noise window in the blanking period and raising the gain threshold to avoid cross-talk (CAG raised to 3/8 of the measured conducted peak in order to diminish sensibility).
In the case of cross-talk, it may be preferable to program a fixed value of atrial blanking post ventricular stimulation to avoid oversensing. When it is not possible to correct the oversensing within acceptable refractory periods, it is possible to modify the atrial sensitivity while trying to respect a sufficient margin in order to detect atrial signals in sinus rhythm and during arrhythmia.