Combined counter - Viva Quad CRT-D

Patient

Patient implanted with a single-chamber defibrillator (Visia AF XT VR) for dilated cardiomyopathy.


Trace

1-2 What diagnosis is suggested by the interval plot?

Two detection zones were programmed: a VT zone from 360 ms and a VF zone from 280 ms; this episode was classified as VF (1) by the device; the interval plot shows a sudden acceleration with cycles detected at the limit of the VF zone; detection of a VF episode effectively treated by a burst (2).

3- What is your diagnosis?
The EGM shows a regular, monomorphic tachycardia, with detection alternating between the VT and VF zones.

4- What therapy is delivered?
The combined counter was full, with 35 cycles classified as TS or FS (7/6 of 30 cycles); analysis of the last 8 cycles before diagnosis showed at least one cycle classified as FV leading to a diagnosis of VF; the first therapy in the VF zone, a burst during charging, was delivered with termination of arrhythmia.

5- What is the diagnosis made by the system?
The diagnosis is VF via the combined counter.

 

Take home message

  • In a MedtronicTM defibrillator , the VT and VF counters operate independently, with a cycle classified as FS not incrementing the VT counter; this differs from the operation of some competitors’ devices and can lead to a delay in diagnosis when an arrhythmia fluctuates between the VT and VF zones, as in this example.
  • The combined counter has been designed to remedy this problem and is systematically active (cannot be programmed off) if a VT zone and a VF zone have been programmed.
  • The combined counter is systematically triggered after the detection of 6 cycles classified as FS (essential prerequisite); the combined counter sums up the cycles classified as TS and FS and is filled when 7/6 (non-modifiable ratio) of the number of cycles required for initial detection in the VF zone have been detected; in this example, this parameter was programmed at 30/40, so the combined counter was filled after 35 cycles classified as TS or FS (7/6 of 30 = 35).
  • When the combined counter is full, the device analyses the last 8 ventricular cycles; if at least one of these cycles is classified as FS, the device diagnoses VF and the first therapy in the VF zone is delivered; if the last 8 cycles are classified as TS, the device diagnoses VT and the first therapy in the VT zone is delivered.