69-year-old man implanted with an Evia DR-T dual-chamber pacemaker for paroxysmal atrioventricular block; during the consultation, automatic right ventricular threshold measurement.
Tracing 20a
Patient: 74-year-old woman implanted with an Evia DR-T dual-chamber pacemaker for complete atrioventricular block; during the consultation, automatic right ventricular threshold measurement.
Tracing 20b
In all modern pacemakers, it is now possible to program an automatic ventricular threshold measurement function more or less associated with an automatic adjustment of the pacing amplitude.
When the ventricular capture control is programmed to OFF, the thresholds are not automatically measured by the device. When the ventricular capture control is programmed to ON, automatic measurement of the ventricular pacing threshold is followed by an adjustment of the pacing amplitude with cycle-to-cycle control of capture efficiency. The principle of the threshold measurement is to determine the minimum amplitude allowing an effective capture with an accuracy of 0.1 V. When the ventricular capture control is programmed to ATM, the device automatically measures the threshold value but without any possible adjustment of the pacing amplitude which remains fixed regardless of the measured value.
The scheduling of the automatic measurement and the frequency of the measurements are programmable.
When the ventricular capture control is programmed to ON, there is therefore a cycle-to-cycle verification of the capture based on the analysis of the evoked response. A safety margin with respect to the threshold measurement is programmable (nominal value 0.5 V). This programming option is only possible if the pulse duration is programmed to 0.4 ms or less; indeed, if a loss of capture is found, a safety pulse is delivered by increasing the pulse duration; hence, a sufficient margin is required.
Another key element to be aware of is that the maximum amplitude that can be delivered is 4.8 V for 0.4 ms, thus well below the maximum capacities. This programming option must therefore be avoided when the pacing threshold is very high.
Various parameters are thus programmable: