Patient with ischemic cardiomyopathy implanted with a dual-chamber ICD (Evera XT DR), hospitalized for lipothymias; history of paroxysmal AF; this tracing illustrates the problem of atrial oversensing by the ventricular lead.
On the graph, there is a regular and fast atrial activity and a railroad track pattern at the level of the ventricular intervals with a very short interval (at the limit of the programmed blanking value) and a longer interval; at the end of the tracing, over a few intervals, very fast ventricular rate superimposed on the atrial activity.
This tracing thus shows an example of atrial activity oversensing by the ventricular canal during an episode of atrial tachycardia. The graph is characteristic of an oversensing of a physiological signal (P-wave, T-wave, R-wave double counting) with a railroad track pattern and an alternation between long and short ventricular intervals. The ventricular EGM shows an alternation between 2 signals of different morphology, the first corresponding obviously to the QRS complex, the second to an oversensing of atrial activity. Oversensing issues are of particular concern in dependent patients since
1) they may cause ventricular pauses,
2) the oversensing may be prolonged in the absence of spontaneous QRS complexes, the level of sensitivity remaining maximal and adapted to the amplitude of the oversensed signals which is often low.
Oversensing of atrial depolarization by the right ventricular lead is rare and occurs mainly in patients implanted with an integrated bipolar lead. In patients in sinus rhythm, the right ventricular lead senses both atrial and ventricular depolarizations, with the PR interval being longer than the ventricular blanking period. If the patient has a complete atrioventricular block, P-wave oversensing may inhibit ventricular pacing and lead to asystole. Similarly, oversensing of atrial depolarization during atrial flutter or atrial tachycardia can cause both inappropriate therapies and asystole if the patient is dependent.
Atrial oversensing by the ventricular lead preferentially occurs in 2 clinical situations:
Atrial oversensing may also be observed under more exceptional circumstances:
Several solutions can be considered to address the problem and avoid the occurrence of inappropriate therapies or ventricular pauses in a dependent patient: