Patient - EN
75-year-old woman implanted with an Adapta pacemaker for complete atrioventricular block post-aortic valve surgery; routine control 3 months after implantation; asymptomatic.
Graph and trace
- programming in AAIÛDDD mode without rate response;
- quasi-permanent ventricular pacing; limited percentage of atrial pacing; this pacemaker-dependent patient presents late PVCs potentially explaining the 1.1% of AS-VS;
- atrial histogram showing a quasi-permanent atrial spontaneous rhythm and a satisfactory rate distribution;
- the ventricles are paced; the ventricular rate curve is modeled on that of the atrial rate.
NID old - EN
3310
As in the previous report, this first page is used to verify that the system is functioning properly (no signs of wear, proper functioning of the 2 leads; the sensing could not be tested, patient appearing pacemaker-dependent). This patient is also programmed in AAIxDDD mode. The histograms, on the other hand, are reversed: