Skip to main content

Dual channel oversensing of what?

Submitted by marc.strik on
Type traces
EGM
Device
PM
Manufacturer
Microport CRM

66-year old patient implanted with a two-chamber pacemaker (Sorin Kora 100 DR) in 2014.

Interrogation revealed 1% ventricular pacing, stable threshold, sensitivity and impedance of both leads.

There was one episode of an atrial mode switch, which occured a few months earlier.

Does anybody have an idea what is causing the oversensing?

Trace PDF

Submitted by marc.strik on Fri 23/12/2016 - 12:06

Permalink

Patient did not recall anything special that day. So no external atrial pacing.

We sent the EGM to Livanova (Paris)..

Submitted by piotrszamlewski on Tue 27/12/2016 - 01:06

Permalink

I would think about placement of electrodes and atrial rhythm origin: Is the ventricular electrode placed properly in RV? It looks like CS signal... but at the same time shows the atrial iEGM a dual signal too. It could be theoretically a double counting of sinus signal/atrial focus due to a two "pathways" between sinusnode/atrial focus and atrial electrode. At the same time a far-field sense of "this second atrial signal" in RV electrode. 

 

 

Submitted by sylvain.ploux on Wed 28/12/2016 - 15:04

Permalink

Hi Marc,

The PM senses correctly the atrial signal around 3mV, and the V signal around 8mV. A supernumerary signals is present in between, on both channels. This signal is not as sharp as the near-field ones.

Physiologically there should be nothing in between the atrial and ventricular activation. I would go for a hardware issue (Minute Ventilation algorithm or so). Let see what Livanova says about it.

Nice brain teaser!

Sylvain