Skip to main content
Défibrillateurs automatiques implantables Medtronic
Masterclass Mort Subite
Microport Academy
DAI Boston Scientific
BIOMONITOR IIIm BIOTRONIK

ECG

ECG, practice reading and et interpreting.

DDD mode

Voir la suite de cet article sur Cursus ECG
Patient
Same patient as previous tracing; device programmed to DDD mode 70 bpm;
Trace
Atrial unipolar pacing (high-amplitude stimulus) and ventricular unipolar pacing after a fixed atrioventricular delay;
Trace
Repeated AP-VP cycles (atrial and ventricular pacing) at 70 bpm;
Comments

The DDD mode is the standard programming mode of dual-chamber pacemakers or resynchronization devices. It ensures atrioventricular synchronization at rest as well as during exercise during sensed or paced atrial activity.

Exergue
DDD mode is designed to respond to the characteristics presented by all implanted patients as a whole but may be associated with a high percentage of deleterious ventricular pacing in patients with preserved atrioventricular conduction.
Haut de page
Tracé
33
Library
Conduction disorders
Pathology
Cardiac pacing
Tags