Atrial sensing failure
Tracing
Manufacturer Biotronik
Device PM
N° 25
Patient
50-year-old man, implanted with an Evia VR pacemaker for sinus dysfunction; recording of a periodic EGM.
Graph and trace
- atrial pacing;
- atrial sensing failure with pacing in the atrial vulnerable period.
Other articles that may be of interest to you
EGM recordings
EGM recordings
EGM recordings
This tracing reveals an intermittent atrial undersensing. P wave sensing has certain peculiarities. Indeed, the amplitude of the sensed atrial signal can vary according to the position of the patient and as a function of the respiratory cycle. It is therefore necessary to set a sufficient margin to avoid atrial undersensing issues during exertion which, in a patient with atrioventricular block, can cause a sudden decrease in the ventricular pacing rate. In addition to the lack of proper monitoring of P waves during exertion, atrial undersensing may have a pro-arrhythmogenic effect if atrial pacing occurs in vulnerable atrial periods with a risk of atrial arrhythmia induction.