Atrial undersensing
Tracing
Manufacturer Medtronic
Device PM
N° 3
Patient
- 69-year-old man
- Dual chamber pacemaker (Medtronic) for complete atrioventricular block following post-surgical aortic valve replacement
- Numerous episodes of atrial arrhythmia in the postoperative phase
- Control 1 month post-implantation
Graph and trace
- P wave not detected by the device
- Atrial pacing at the end of the escape period; this pacing occurs in the vulnerable period of the previous undetected P wave; ventricular pacing
- Intermittent atrial undersensing
Comments
- This tracing reveals an intermittent atrial undersensing
- Sensing of the P wave presents certain specificities, the slope of the signal and its amplitude being lower than those of a QRS complex
- The amplitude of the sensed atrial signal may vary according to the position of the patient and as a function of the respiratory cycle
- Atrial undersensing on exertion in a patient with atrioventricular block may cause a sudden drop in ventricular pacing rate
- It is therefore necessary to set a sufficient margin to avoid atrial undersensing on exertion
- Atrial undersensing may have a pro-arrhythmogenic effect if atrial pacing occurs in an atrial vulnerable period with a risk of AF induction