Device-related alert message: de-activated ventricular classification
Tracing
Manufacturer Biotronik
Device Remote monitoring
Field Messages and events
N° 4
Patient
This 43-year-old man received a Lumax 340 VR-T single chamber defibrillator for hypertrophic cardiomyopathy with episodes of sustained ventricular tachycardia. He underwent elective surgery on a salivary gland.
Graph and trace
Telemedicine report
Within 24 h after his return home, a systematic alert message was transmitted warning that the detection arrhythmias had been turned ‘off’ (red color status).
Other articles that may be of interest to you
This patient underwent elective surgery, in which case 2 choices were available:
In this case, the detection was turned off before surgery, and the defibrillator was neither interrogated nor reprogrammed at the end of the operation. This could have had catastrophic consequences, had a serious arrhythmia developed. Telemedicine enabled the early diagnosis of this management error, the patient was called in the next day, and the detection and therapies were programmed back ‘on’. This example illustrates one of the advantages of remote follow-up. This patient would not have been protected by his defibrillator for several months had he been seen exclusively for standard face-to-face ambulatory visits.