Dual chamber discrimination, threshold for correlation
Tracing
Manufacturer Boston Scientific
Device ICD
Field Discrimination
N° 11
Patient
This 58-year-old woman underwent implantation of an Incepta single chamber defibrillator.
Graph and trace
- sinus tachycardia;
- acceleration in the slow VT-1 zone (520 ms);
- after 3 cycles classified VT-1, analysis of the correlation;
- 8 out of 10 criterion fulfilled;
at the end of Duration, diagnosis of SVT based on high correlation between morphology during tachycardia and morphology of reference EGM; therapy withheld;
- continuation of cycle-by-cycle analysis;
- on this cycle, the criterion of correlation was no longer fulfilled. The analysis reveals <3 out of 10 cycles with >94% correlation (a programmable threshold with this type of device), despite minimal variations in the EGM morphology on the high-voltage channel. The decision to treat was made ;
- burst of ATP;
- persistence of sinus tachycardia; during redetection, no discrimination (no analysis of the level of correlation);
- second burst of ATP;
- third burst of ATP;
- the ventricular rate slows slightly and falls below the VT zone;
Other articles that may be of interest to you
EGM recordings
This episode highlights the importance of the possibility of programming the threshold for correlation. This episode of sinus tachycardia is initially appropriately diagnosed as a supra-ventricular tachycardia. However, the percentage of correlation hoovers around the programmed threshold of 94%. During a certain number of cycles, a majority of the measurements are 93% (below the threshold) explaining the occurrence of inappropriate therapies. The threshold was therefore modified to a lower value solving the problem.