Lead fracture and multiple charges of the capacitors

Tracing
N° 4
Manufacturer Boston Scientific Device ICD
Patient

This 73-year-old man underwent implantation of a Boston Science dual chamber defibrillator 8 years earlier, for secondary prevention of sustained VT in the context of ischemic cardiomyopathy. He underwent replacement of the pulse generator and implantation of a Teligen defibrillator. Multiples episodes of VT and non-sustained VF, requiring 1 or 2 sequences of ATP, were observed during a routine ambulatory visit. 

Graph and trace

Tracing

  1. AP-Sr VP-Sr cycle, where Sr means sensor; therefore, the rate was set by the sensor: sensor-responsive mode;
     
  2. oversensing of signals of variable amplitudes and morphologies by the ventricular channel in the VT or VF zones; the high-voltage channel was free from oversensing, allowing the differentiation of the spontaneous ventricular events; the atria were paced at the sensor-driven rate; the atrial rate was variable as it was influenced by ventricular oversensing;
     
  3. VF counter fulfilled (V-Epsd);
     
  4. intermittent oversensing; at the end of Duration, the rhythm was diagnosed as sinus; no therapy was delivered;
     
  5. further ventricular oversensing;
     
  6. the VF counter was fulfilled and, at the end of Duration, the first therapy was delivered;
     
  7. during ATP (Quick Convert), ventricular capture was intermittent (2 effective and 6 ineffective stimuli);
     
  8. oversensing continued after ATP and the charge of the capacitors begins (the first cycle after the ATP sequence was not counted; thereafter, 2 out of 3 cycles were classified in the VT or VF zone, explaining the onset of the capacitors charge);
     
  9. at the end of oversensing, the charge was aborted (the 6 out of 10 short cycles criterion was no longer fulfilled);
     
  10. oversensing resumed;
     
  11. the VT counter was fulfilled (V-Detect);
     
  12. a sequence of ATP was delivered; the ventricular stimuli failed to capture.

Graph

The graph suggests a lead dysfunction with presence of a cloud of ventricular events inconsistent with a ventricular arrhythmia.

Comments

The association of an abnormal impedance and a tracing showing intermittently short and disorganized cycles of variable amplitude, was highly suggestive of lead dysfunction. The graph also showed a characteristic aspect of a cloud of points with very short cycles at the limit of the ventricular blanking period. In this asymptomatic patient, remote monitoring would have enabled an early diagnosis, which could have prevented the delivery of inappropriate therapies. Likewise, an early diagnosis would have prevented a succession of inappropriate charges, which, even when aborted, consume energy and can markedly shorten the lifespan of the device. 

Message to remember

Detection and impedance curves

Starting in July 2011, the impedance curve of atrial pacing showed a clear increase and several >2,000 Ohms measurements. The RV shock impedance remained relatively stable. However, several measurements of ventricular pacing impedance <200 Ohms were observed during the month of July.

Sequence of events since the last interrogation

The review of the sequence of events revealed the presence of multiple episodes classified in the VT zone, or of sustained or non-sustained VF, sometimes requiring 1 or 2 sequences of ATP. The sudden occurrence of these arrhythmias was approximately simultaneous with the recording of abnormal measurements of ventricular pacing impedance.

Summary of episode

In a first stage, the episode was classified in the VF zone. A sequence of ATP was delivered, though the shock was aborted because of failure of reconfirmation; a second episode was diagnosed in the VT zone; a salvo of ATP was delivered.